Reliability, Learnability and Efficiency of Two Tools for Cement Crowns Retrieval in Dentistry



Cristina Bignardi1, *, Elisabetta M. Zanetti2, Mara Terzini1, Anna R. Ciccola1, Gianmario Schierano3, Alberto L. Audenino1
1 Department of Mechanical and Aerospace Engineering, Polytechnic University of Turin, Turin, Italy
2 Departiment of Engineering, University of Perugia, Perugia, Italy
3 Department of Surgical Science, Dental School, C.I.R University of Turin, Turin, Italy


Article Metrics

CrossRef Citations:
4
Total Statistics:

Full-Text HTML Views: 991
Abstract HTML Views: 593
PDF Downloads: 235
ePub Downloads: 223
Total Views/Downloads: 2042
Unique Statistics:

Full-Text HTML Views: 518
Abstract HTML Views: 374
PDF Downloads: 172
ePub Downloads: 169
Total Views/Downloads: 1233



Creative Commons License
© 2018 Bignardi et al.

open-access license: This is an open access article distributed under the terms of the Creative Commons Attribution 4.0 International Public License (CC-BY 4.0), a copy of which is available at: https://creativecommons.org/licenses/by/4.0/legalcode. This license permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.

* Address correspondence to this author at the Department of Mechanical and Aerospace Engineering, Polytechnic University of Turin, Cso Duca degli Abruzzi 24, 10129 Turin, Italy; Tel: +39-011-0906944; Email: cristina.bignardi@polito.it


Abstract

Background:

Tooth-supported fixed dentures are commonly used in restorative dentistry, and have definitely reached a high survival rate; nevertheless, their removal is sometimes required mainly due to caries or other failures (poor fit, poor cementation and so on). Removing a definitive partial denture is not trivial since the used cement is not always known and it may be very resistant; additionally, there are various clinical circumstances in which a conservative disassembly would be desirable.

Objective:

Assessing the performance of different tools for cement crowns retrieval in terms of reliability, learnability and efficiency.

Methods:

An experimental study has been performed on two different devices for conservative crown/bridge removal: a manual tool that is a sliding hammer, and an automatic tool, powered by compressed air. Both skilled and unexperienced operators have been considered and an experimental set up has been appositely designed in order to measure force versus time patterns. The peak applied force has been taken as an output variable for the evaluation of tool performance.

Results:

The automatic tool improves both the inter-operator and the intra-operator reliability, respectively from 79% to 95%, and from 69% to 92%. Additionally, the force pattern is significantly different between these two tools: the instrument powered by compressed air, produces a sharper peak force, as required to break fragile materials such as dental crown cement, and its efficiency can be estimated to be 75% higher. Both tools have a high learnability since the performances of experienced and unexperienced operators have not proved to be significantly different.

Conclusion:

A methodology has been set up to compare tools for cement crowns retrieval. The compressed-air tool has been proved to generally provide a better performance unless more ductile cement is to be broken.

Keywords: Retrieval, Cement, Compressed air, Inverted hammer, Coronaflex, Crown, Bridge, Removal.