Table 3: Commercially available POC devices for diabetes management.

POC NAME MEASURED PARAMETERS TEST TIME IMPRECISION ACCURACY
StatStrip® (Nova Biomedical, Waltham,
MA, USA) (119-123)
glucose 6 seconds CV = 5 - 7% r = 0.99
NovaStatstrip® (Nova Biomedical, Waltham, MA, USA) (124-126) glucose < 10 minutes CV = 3.0 - 4.7% r = 0.97 - 0.98
Precision Xtra™ (MediSense/Abbott Diabetes Care, Abbott Park, IL) (127, 128) BHB 10 seconds CV = 2.4 - 5.9% r = 0.92
Accu-Chek Inform® II (Roche Diagnostics, Belgium) (129) glucose 5 seconds CV = 1.7 - 3.7% r = 0.99
Precision Xceed Pro blood glucose® (Abbott, USA) (128, 130] BHB 20 seconds CV = 2.4 - 5.9% r = 0.99
Assure Platinum blood glucose monitoring system® (Arkray, USA) glucose 7 seconds CV = 1.9 - 4.9% r = 0.99
HemoCue® Glucose 201+ System (HemoCue®, Sweden) [131-133] glucose 1 minute CV = 2.52 - 3.66% r = 0.96
YSI 2300 STAT® Plus (YSI Life Sciences, USA) [134-139] glucose and lactate 45-65 seconds CV = 3.03% r = 0.96
Scan SureStep Flexx (LifeScan, Johnson & Johnson Company, USA) [140] glucose 15 seconds CV = 1.4 - 4.1% r = 0.98
NycoCard II (Axis-Shield, Norway) (141-143) HbA1c 3 minutes CV = 3.1% r = 0.93
Accu-chek Advantage II (Roche, Basel, Switzerland) [144, 145] glucose 5 seconds CV = 0.07 - 2.5% r = 0.99
Precision PCx (Abbott, Illinois, USA) [144, 146] glucose 20 seconds CV = 3.8 -5.4% r = 0.96
SureStep Flexx (LifeScan, Malpitas, CA) [147] glucose 15 seconds CV = 4.2 - 5.9% r = 0.68
DCA VantageTM (Siemens Healthcare Diagnostics Inc, Australia) [148-150] Albumin/creatinine, HbA1c, glucose 6 minutes CV = 1.7 - 3.6% r = 0.74 - 0.98
FreeStyle Precision Neo (Abbott, Diabetes Care Ltd, UK) [151] glucose, BHB 5 seconds CV < 3.5% r = 0.89
StatStrip® (Nova Biomedical, Waltham,
MA, USA) [119-123]
glucose 6 seconds CV = 5 - 7% r = 0.99
NovaStatstrip® (Nova Biomedical, Waltham, MA, USA) [124-126] glucose < 10 minutes CV = 3.0 - 4.7% r = 0.97 - 0.98
LABGEO PT10 (Samsung Electronics, Suwon, South Korea) [152] HbA1c 7 minutes CV = 2.6% r = 0.99
A1C EZ 2.0 (Biohermes,
Wuxi, China) [153]
HbA1c 3 minutes CV = 1.9 - 2.3% r = 1.00
DPN-Check (Neurometrix Inc., Waltham, MA) [154] nerve amplitude potential and sural nerve conduction velocity 2 minutes CV = 3.6 - 8.8% r = 0.67
FIA8000 (GeTein BioMedical Inc., Portland, USA) [102] HbA1c 10-20 minutes CV = 1% r = 0.99
FastPack® IP System (Sekisui Diagnostics, LLC, Lexington, USA) [114] TSH 12 minutes CV = 4 - 7% r = 0.97
vitamin d CV = 4.7 - 15.1% r = 0.92
free T4 7 minutes CV = 7.2 - 11.5% r = 0.95
Clini5
(Callegari S.r.l., Parma, Italy) [7]
Glu 10 - 180 seconds CV = 1.37 - 5.38% r = 0.75 - 0.99
HbA1c 5 - 10 minutes

(BHB = β-hydroxybutyrate; TSH = thyroid stimulating hormone; T4 = Thyroxine, HbA1c = glycated hemoglobin)