POC NAME | MEASURED PARAMETERS | TEST TIME | IMPRECISION | ACCURACY |
---|---|---|---|---|
StatStrip® (Nova Biomedical, Waltham, MA, USA) (119-123) |
glucose | 6 seconds | CV = 5 - 7% | r = 0.99 |
NovaStatstrip® (Nova Biomedical, Waltham, MA, USA) (124-126) | glucose | < 10 minutes | CV = 3.0 - 4.7% | r = 0.97 - 0.98 |
Precision Xtra™ (MediSense/Abbott Diabetes Care, Abbott Park, IL) (127, 128) | BHB | 10 seconds | CV = 2.4 - 5.9% | r = 0.92 |
Accu-Chek Inform® II (Roche Diagnostics, Belgium) (129) | glucose | 5 seconds | CV = 1.7 - 3.7% | r = 0.99 |
Precision Xceed Pro blood glucose® (Abbott, USA) (128, 130] | BHB | 20 seconds | CV = 2.4 - 5.9% | r = 0.99 |
Assure Platinum blood glucose monitoring system® (Arkray, USA) | glucose | 7 seconds | CV = 1.9 - 4.9% | r = 0.99 |
HemoCue® Glucose 201+ System (HemoCue®, Sweden) [131-133] | glucose | 1 minute | CV = 2.52 - 3.66% | r = 0.96 |
YSI 2300 STAT® Plus (YSI Life Sciences, USA) [134-139] | glucose and lactate | 45-65 seconds | CV = 3.03% | r = 0.96 |
Scan SureStep Flexx (LifeScan, Johnson & Johnson Company, USA) [140] | glucose | 15 seconds | CV = 1.4 - 4.1% | r = 0.98 |
NycoCard II (Axis-Shield, Norway) (141-143) | HbA1c | 3 minutes | CV = 3.1% | r = 0.93 |
Accu-chek Advantage II (Roche, Basel, Switzerland) [144, 145] | glucose | 5 seconds | CV = 0.07 - 2.5% | r = 0.99 |
Precision PCx (Abbott, Illinois, USA) [144, 146] | glucose | 20 seconds | CV = 3.8 -5.4% | r = 0.96 |
SureStep Flexx (LifeScan, Malpitas, CA) [147] | glucose | 15 seconds | CV = 4.2 - 5.9% | r = 0.68 |
DCA VantageTM (Siemens Healthcare Diagnostics Inc, Australia) [148-150] | Albumin/creatinine, HbA1c, glucose | 6 minutes | CV = 1.7 - 3.6% | r = 0.74 - 0.98 |
FreeStyle Precision Neo (Abbott, Diabetes Care Ltd, UK) [151] | glucose, BHB | 5 seconds | CV < 3.5% | r = 0.89 |
StatStrip® (Nova Biomedical, Waltham, MA, USA) [119-123] |
glucose | 6 seconds | CV = 5 - 7% | r = 0.99 |
NovaStatstrip® (Nova Biomedical, Waltham, MA, USA) [124-126] | glucose | < 10 minutes | CV = 3.0 - 4.7% | r = 0.97 - 0.98 |
LABGEO PT10 (Samsung Electronics, Suwon, South Korea) [152] | HbA1c | 7 minutes | CV = 2.6% | r = 0.99 |
A1C EZ 2.0 (Biohermes, Wuxi, China) [153] |
HbA1c | 3 minutes | CV = 1.9 - 2.3% | r = 1.00 |
DPN-Check (Neurometrix Inc., Waltham, MA) [154] | nerve amplitude potential and sural nerve conduction velocity | 2 minutes | CV = 3.6 - 8.8% | r = 0.67 |
FIA8000 (GeTein BioMedical Inc., Portland, USA) [102] | HbA1c | 10-20 minutes | CV = 1% | r = 0.99 |
FastPack® IP System (Sekisui Diagnostics, LLC, Lexington, USA) [114] | TSH | 12 minutes | CV = 4 - 7% | r = 0.97 |
vitamin d | CV = 4.7 - 15.1% | r = 0.92 | ||
free T4 | 7 minutes | CV = 7.2 - 11.5% | r = 0.95 | |
Clini5 (Callegari S.r.l., Parma, Italy) [7] |
Glu | 10 - 180 seconds | CV = 1.37 - 5.38% | r = 0.75 - 0.99 |
HbA1c | 5 - 10 minutes |